Measurement range -0 to 1.5 ppm Al (0 to 1.5 mg/L)
Chemical Method- Pyrocatechol Violet (PCV)
Background color correction Compensated at the measurement wavelength
Measurement frequency-1 to 6 measurements per hour (User-selectable)
Accuracy / maximum measured error ±5 % of reading or ±0.005 ppm (whichever is the greater)
Repeatability ±5 % of reading or ±0.005 ppm (whichever is the greater)
Resolution-0.001 ppm or 1 ppb
Measurement units mg/l, ppm, ppb, μg/l
Calibration2-point, automatic calibration
Measurement range-0.050 to 1000 ppm NH3 (0.050 to 1000 mg/l)
Measurement Principle Gas-sensing ammonia probe
Measurement mode-Continuous
Accuracy / maximum measured error,±5 % of reading or ±0.02 ppm for 0.05 to 500 ppm (whichever is the greater),±7.5 % of reading for 500 to 1000 ppm
Repeatability,±2 % of reading or ±0.02 ppm (whichever is the greater)
Calibration 2-point, automatic calibration
Measurement units mg/l, ppm, ppb, μg/l
Total and free residual chlorine measurement – add potassium iodide to reagent to convert free chlorine measurement to total chlorine
Reagentless operation – reduced operating costs with reagentless operation at a pH of 7.5 or less
Multiple sensor input capability – reduced installation costs for multiple point monitoring operations
Self-cleaning sensor assembly – prolongs sensor life and maintains accuracy
Stainless steel electrodes enabling measurements down to 0.055 µS/cm ultra-pure water
Enables installation without the need for costly bypasses
Enables maintenance without process shutdown
1 in. BSP or NPT threaded process connection
Guaranteed 0.05 and 0.10 cell constants
Integral Pt100 temperature sensor
Maximum temperature 110 °C (230 °F)
Maximum pressure 10.5 bar (150 psi)
Integral 0.9 m (3 ft) cable
Transmitter options for all sensor types
— 2-electrode sensors for low conductivity measurement
— 4-electrode sensors for wide range conductivity applications and slurries
— Toroidal conductivity for corrosive and heavy coating applications
Through-the-glass programming (TTG) configuration
Hazardous area approved- IS/NI and Exd
Communications- 4 to 20 mA with HART (FDT/DTM/EDD)
Continuous sensor and self-monitoring- Electronics self-check
Annular carbon electrodes enabling measurements up to 10,000 µS/cm
Screw-in insertion sensor with 1 in. BSP or NPT threads
0.10 and 1.0 cell constants
Integral Pt100 temperature sensor
Maximum temperature 100 °C (212 °F)
Maximum pressure 7 bar (100 psi)
Ingress protection IP 68
Fixed cable, terminal head or detachable cable versions
Reduced sensor fouling design
Low maintenance, long-life sensor
Complete systems and kit versions available
On-line water wash option
Reagentless operation
Automatic cleaning
Virtually zero maintenance
Dual-wavelength measurement for turbidity and dissolved organic compensation
Long lamp life
Simple robust design in three dip lengths
Positive salt bridge flow from integral reservoir
Easily-replaced reliable combination electrode
Built-in sight glass for salt bridge reservoir
Simple installation on wall or tank
User-customizable for insertion, f-through, immersion and hot-tap retractor
Economical investment, powerful performance
Combination Sensor complete with measuring, reference and temperature elements
Bulb and flat glass options
High stability PTFE reference junction
Gel-filled disposable sensor
Process-resistant gel electrolyte
The Multiwave Photometer PFO3372 incorporates the use of fiber optic sampling accessories for added measurement capability.
The ABB Multiwave photometers provide continuous measurements in complex streams which allow for fast reactions to process changes.
The fast analysis provides optimum control of a process and improves its safe operation.
Measurements can be made in liquid or vapor streams. Simple installation on wall or tank.
Oxygen only or oxygen plus combustibles
– increased combustion efficiency
– burner malfunction identification
– enhanced plant safety
Comprehensive diagnostics
– NAMUR-compliant diagnostic symbols
– supports predictive maintenance
– fully logged diagnostic events
Process logging and trending of all measured and
calculated values
– oxygen and carbon monoxide equivalent (COe)
– process temperature measurement
– combustion efficiency calculation
Multiple sample filter and blowback options, optional dual filtration system, fully programmable blowback.
Real-time continuous on-line measurements of several gases simultaneously
Gases measured include CH4, C2H2, CO, CO2, H2O, H2S, HCl, HF, NH3, O2, and others
User selectable data rate
Industry Standard Communications, including analog 4 to 20 mA and digital Modbus
Hazloc-approved USB portal access for data file download and simple field service
Compact footprint – 16 inches, front to back, 28 pounds
Conventional analytical components, not a GC on a chip
Windows CE® operating system
Interactive, graphical display actuated by magnet contact
Range -Turbidity: 0 to 4000 NTU,Suspended solids: 0 to 100,000 mg/l SiO2 (100 g/l)
Accuracy-Turbidity: <±2 % measured value,Suspended solids: dependent on sample
Method-90° scattered light measurement in accordance with DIN/EN 27027/ISO 7027
Repeatability,<1 %
Limit of detection-0.006 NTU
Response time-T90 < 30 s with filtering disabled
Storage conditions–5 to 70 °C (23 to 158 °F)
Operating temperature-0 to 60 °C (32 to 140 °F)
Operating pressure-Up to 10 bar (145 psi) for metal versions